Inductive Clip-On Extensometers

These extensometers are attached directly to specimens and are designed for extension measurement on metals and plastics.

Their range of application covers all fine strain and offset yield measurement, together with determination of Young's modulus. Clip-on extensometers are available with manual or automatic attachment.

Key Advantages and Features

Advantages and features

The inductive clip-on extensometer is intended for fixed measurement lines L0 = 10 mm and L0 = 20 mm (optionally up to L0 = 100 mm) with a measuring range of 2 mm in the tensile and compression directions.

  • As the measuring system is located very close to the specimen, the measurement lever-arms involved are very small, giving increased accuracy and reproducibility.
  • Very high resolution over the entire measurement range.
  • Fitted with two measuring systems for dual averaging measurement of outer fiber strain.
  • Suitable for tensile, compression and cyclic load tests.

Weight Relieving option

The inductive clip-on extensometer can optionally be equipped with a with weight relieving device.

Use of the stabilizer arm allows the clip-on extensometer to be attached more easily and reproducibly. The weight on the specimen is also relieved so that no torque is applied, allowing the clip-on extensometer to be used for thin and sensitive specimens also. The extensometer is additionally protected against slipping and falling (in the event of the specimen breaking or being changed).

  • Advantages of this option
  • Extensometer weight relieving for sensitive specimens (force adjustable).
  • Can be swiveled out for specimen change.

Technical data

Extensometer with 2 inductive systems for AllroundLine and zwickiLine testing systems

Item No.

056896

058008[1]

Item No.

029443

Initial gage length

20, can be reset to 10

20, can be reset to 10

mm

Measurement travel

2

2

mm

Measured-value resolution

depends on machine electronics, see machine electronics resolution

Depends on machine electronics, see machine electronics resolution

Measurement signal

totaled

Two separate signals

Accuracy

Class 0.5

Class 0.5

To EN ISO 9513

Specimen dimensions:

Round specimens, max.

Ø40

Ø40

mm

Flat specimens, max.

40 x 40

40 x 40

mm

Round specimens, max.

Flat specimens, max.

  1. 2 x USC module required

Extensometer (Inductive) for Measurement of Fine Strain


Technical data for fine strain measurement

Extensometer (inductive) for fine-strain measurement in conjunction with multiXtens or makroXtens II.
For change-of-width measurement use only with 15° swiveling transverse strain extensometer and 1 or 2 measuring lines (Item nos. 057088, 057089) for multiXtens or 084637, 084638 for makroXtens II.

Item No.

056895

058001[1]

Initial gage length

20 to 50

20 to 50

mm

Measurement travel

2

2

mm

Accuracy

Class 0.5

Class 0.5

To EN ISO 9513

  1. 2x USC module required (two separate signals)